Main Technical Specifications
Power Supply Voltage Single Phase AC 180~240V
Frequency 50 Hz
Current (Max., instantaneous) <20A
Power capacity greater than 5.5kvA
Internal resistance of power supply wire ≤1Ω
Maximum rated capacity Current(mA) Voltage (kvp) Exposure time(s)
30 90 0.08~6.3
40 90 0.08~4.0
50 90 0.08~1.0
80 0.08~2.5
Timer 0.08s to 6.3s, in 20 shifts, electronic type
Max distance from tube focus to the ground ≥1820 mm
Min distance from tube focus to the ground ≤550 mm
Remote control exposure distance ≥ 6m
Max. exposure field 350×350 mm (when distance between X-ray tube focus and film is 650 mm)
Max. exposure field 450×450 mm / when distance 1000mm
Specification of x ray tube XD1-3/100, stationary anode, focus 2.3 x 2.3